Analytical performance data

Analytical performance was evaluated on PATHFAST system with whole blood and plasma
[Okamura Y, Ref. 6; Internal data]. 

Assay range 20 – 20,000 pg/mL
Correlation between
whole blood and
plasma on PATHFAST
y = 1.04 x - 10.8 ; r = 0.986 ; n = 104
(y: EDTA whole blood, x: EDTA plasma)
 
Total % CV
in plasma
QC-LL= 4.4%, QC-L= 4.0%, QC-M = 3.8%,
QC-H = 5.0%
Turnaround time 15 minutes
Sample material
 
anticoagulated (EDTA/Heparin)
whole blood or plasma